Depth Sensitivity and Source-Detector Separations for
Near Infrared Spectroscopy Based on the Colin27 Brain
Template
Gary E. Strangman*, Zhi Li, Quan Zhang
Neural Systems Group, Massachusetts General Hospital/Harvard Medical School, Charlestown, Massachusetts, United States of America
Abstract
Understanding the spatial and depth sensitivity of non-invasive near-infrared spectroscopy (NIRS) measurements to brain
tissue–i.e., near-infrared neuromonitoring (NIN) is essential for designing experiments as well as interpreting research
findings. However, a thorough characterization of such sensitivity in realistic head models has remained unavailable. In this
study, we conducted 3,555 Monte Carlo (MC) simulations to densely cover the scalp of a well-characterized, adult male
template brain (Colin27). We sought to evaluate: (i) the spatial sensitivity profile of NIRS to brain tissue as a function of
source-detector separation, (ii) the NIRS sensitivity to brain tissue as a function of depth in this realistic and complex head
model, and (iii) the effect of NIRS instrument sensitivity on detecting brain activation. We found that increasing the source-
detector (SD) separation from 20 to 65 mm provides monotonic increases in sensitivity to brain tissue. For every 10 mm
increase in SD separation (up to ,45 mm), sensitivity to gray matter increased an additional 4%. Our analyses also
demonstrate that sensitivity in depth (S) decreases exponentially, with a ‘‘rule-of-thumb’’ formula S = 0.75*0.85
depth
. Thus,
while the depth sensitivity of NIRS is not strictly limited, NIN signals in adult humans are strongly biased towards the
outermost 10–15 mm of intracranial space. These general results, along with the detailed quantitation of sensitivity
estimates around the head, can provide detailed guidance for interpreting the likely sources of NIRS signals, as well as help
NIRS investigators design and plan better NIRS experiments, head probes and instruments.
Citation: Strangman GE, Li Z, Zhang Q (2013) Depth Sensitivity and Source-Detector Separations for Near Infrared Spectroscopy Based on the Colin27 Brain
Template. PLoS ONE 8(8): e66319. doi:10.1371/journal.pone.0066319
Editor: Xi-Nian Zuo, Institute of Psychology, Chinese Academy of Sciences, China
Received March 21, 2013; Accepted April 29, 2013; Published August 1, 2013
Copyright: ß 2013 Strangman et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits
unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Funding: This work was supported by the National Space Biomedical Research Institute through NASA NCC 9-58. The funder had no role in study design, data
collection and analysis, decision to publish, or preparation of the manuscript.
Competing Interests: The authors have declared that no competing interests exist.
* E-mail: [email protected].harvard.edu
Introduction
Near-infrared spectroscopy (NIRS) and diffuse optical imaging
(DOI) have been successfully used for non-invasive assessment of
brain hemodynamics for over three decades [1,2,3]. Properly
interpreting any type of non-invasive measurement including
NIRS and DOI requires detailed knowledge about the
measurement’s spatial sensitivity profile. In general, the question
is what tissues are being probed by a given measurement? A more
specific question commonly asked is, what is the depth penetration
of NIN? Addressing these questions is particularly difficult because
light propagation through scattering media with heterogeneous
structure (such as the head) is inherently complex, and because the
governing mathematical models of this process the radiative
transport equation and its diffusion approximation are difficult
to solve analytically for any but the most trivial of tissue geometries
[4,5,6,7]. It is even more difficult to empirically measure light
fluence within tissue, as this would require the placement of
omnidirectional light detectors inside a probed medium.
For simple tissue geometries, such as optically homogeneous
tissues with infinite, semi-infinite or slab boundary conditions,
analytical solutions to the diffusion equation have been developed
[8,9,10,11,12,13]. However, even modestly more complex geom-
etries, and especially the irregular boundaries and undulating
layers found in brain tissue, have no analytical solutions. As a
result, far less is understood about the photon distribution through
the head and, consequently, NIRS sensitivity to brain tissue.
In the absence of direct observation or analytical solutions,
questions about sensitivity and penetration depth in complex
tissues must instead rely on numerical approaches. There are two
general categories: (1) approaches based on finite element (FE) and
finite difference (FD) analysis [14,15], or (2) Monte Carlo
simulations of photon propagation through the tissue in question
[16]. The appeal of FE and FD techniques is that they can handle
arbitrary tissue type boundary conditions yet require considerably
less computation time than Monte Carlo methods [15]. A
shortcoming of FE/FD techniques is that they need to assume
an analytical form for the photon migration through tissue.
Typically (though not always, e.g. [17]), the diffusion approxima-
tion is assumed. While the diffusion approximation is generally
well supported, it is difficult to test in detail the required
assumptions, particularly for complex tissue boundary conditions
such as the undulating layers inside a human head.
The other numerical technique typically employed is Monte
Carlo (MC) simulation [16,18,19,20]. This technique proceeds as
follows: (1) select a three-dimensional tissue geometry and divide it
into voxels of different tissue types, (2) assign scattering and
absorption optical properties to each voxel based on tissue type, (3)
select a point on the surface of this geometry and ‘‘inject’’ a photon
PLOS ONE | www.plosone.org 1 August 2013 | Volume 8 | Issue 8 | e66319
at that point, (4) propagate that photon through the tissue by
allowing it to probabilistically scatter and be absorbed as it travels,
and (5) repeat steps 3–4 many thousands or millions of times and
accumulate the resulting photon weights (or, fluences) along with
the cumulative distance traveled through each tissue type. The
Monte Carlo approach is unfortunately quite computationally
intensive. However, it provides the most accurate estimate of
photon propagation through irregular boundary conditions and
heterogeneous media such as the human brain. Importantly, it
does not need to assume an analytical form for photon
propagation, and hence is not biased by diffusion or other
approximations [21,22].
A number of groups have used Monte Carlo methods to
investigate NIRS sensitivity in layered tissues. Most such studies,
however, have only considered relatively simple geometries such as
layered slabs [23,24,25,26,27], cylinders [28,29,30], or concentric
spheres [31]. In general, these studies confirm general expectations
based on diffusion theory: (1) depth sensitivity increases as source-
detector separation increases [31,32], (2) overlying tissue layers
interfere with the measurement of deeper layers [30,33,34], (3)
sensitivity to deeper layers decreases with increasing thickness of
overlying tissue layers [24,25], (4) depth sensitivity is affected by
the angle at which incident photons contact the surface [28], and
(5) there is an improvement of NIRS sensitivity to deeper tissue
layers when modeling scattering elements are embedded in
cerebrospinal fluid (CSF) layers [26]. These studies have provided
very useful guides, but the use of simple geometries means that the
results cannot be assumed to apply to brain geometries, at least not
quantitatively.
An important set of studies focused more specifically on the role
that the relatively clear CSF layer may play in NIN measurements
[23,24,25,26,29,35,36,37,38]. For example, at least one study
suggested that CSF could distort the sensitivity profile, both
broadening it and reducing the sensitivity to brain with increasing
CSF thicknesses from 0.5 to 5 mm [25]. The CSF effect also
appears strongly modulated by the precise amount of scattering
within the CSF layer itself [24,26,35,36]. These CSF-related
effects appear to be substantial relative to other tissue distribution
issues [38].
Six studies have examined realistic human brain models based
on structural MRI scans [32,34,35,38,39,40]. All of these studies
employed a handful of Monte Carlo simulations, focusing on a
particular location of the head. NIRS sensitivity to functional
brain activation was concluded to be ‘‘high’’ [34], at least for
source-detector (SD) separations in the 30–35 mm range [32,39].
In an MC study on two separate MRI-based head models of pre-
term infants, using 1 source plus 20 detectors around the head,
differences in tissue distributions were found to affect measured
NIRS signals, but these effects were modest compared to the
effects generated by the relatively clear CSF [38]. Another study
examined whether the CSF effect was sufficient to invalidate the
use of the diffusion approximation for computing photon
propagation through head tissue [40]. Using 1 source and 25
linearly aligned detectors in an MRI-based adult head phantom,
sensitivity to the brain reached approximately 11% of the
sensitivity profile at source-detector (SD) separations of 40 mm.
This sensitivity was slightly reduced in the face of larger scattering
coefficients, but the overall conclusion was that a relatively
scattering CSF layer (m
s
~0:3mm
{1
) was unlikely to generate
more than 20% error in sensitivity estimates when using the
diffusion approximation.
Most recently, Mansouri and colleagues combined an MRI-
based head model and a highly resource-intensive MC approach
to assess the sensitivity of NIRS to brain tissue over one location in
the left frontal pole [35]. Using four simulations, and examining
source-detector (SD) separations of 20, 33 and 40 mm, they found
sensitivity to brain tissue ranged from approximately 1% of the
total NIRS sensitivity (99% coming from scalp and skull) at a
20 mm SD separation, to 6–9% of the total NIRS sensitivity at
40 mm. The maximum sensitivity again depended on the amount
of scattering in the CSF layer, with higher scattering
(m
s
~1:0mm
{1
) leading to greater brain sensitivity as found by
Custo and colleagues.
The above findings–particularly when coupled to the known
anatomical variability for scalp, skull, CSF, gray and white matter
within and between people and brain regions suggest that small
numbers of MC runs are likely to be biased by regional details in
tissue types and layer thicknesses. The available literature,
however, leaves a notable gap in terms of a more comprehensive
analysis of photon propagation across realistic head geometries.
First, the current maximum of less than 10 Monte Carlo runs in a
single head model is insufficient for comprehensively and robustly
evaluating NIRS sensitivity to complex tissues such as the brain,
much less providing variability measures of such sensitivity.
Second, there has been limited quantification of NIRS sensitivity
to brain versus non-brain tissue compartments as a function of
source-detector separation, or in relation to instrument sensitivity.
And third, NIRS sensitivity as a function of depth remains
particularly poorly understood for complex head geometries, yet
this is key information when trying to interpret NIN data. Here we
address each of these issues using over 3,555 Monte Carlo
simulations in a detailed, five-layer model of the standard Colin27
human head template.
Materials and Methods
Brain Tissue Model
For a well-characterized starting point, we used the 16161mm
Colin27 template brain [41], as distributed with FSL v4.1 [42]. To
this scan, we applied the default SPM8 segmentation process [43]
which generated three tissue type probability images: gray matter,
white matter and cerebrospinal fluid (CSF). Each probability map
was slightly smoothed (0.75 mm FWHM Gaussian kernel) and
then intracranial voxels were classified as gray matter, white
matter or CSF. A voxel was classified as gray matter if it had the
highest probability of being gray matter and this probability was
greater than 20%. White matter was treated similarly. An
intracranial voxel was then classified as CSF if it had a higher
probability than both gray and white matter. Holes were filled
with the modal nearest-neighbor tissue type (less than 0.02% of all
voxels). The original Colin27 template was also passed through
FSL’s brain extraction tool [44] to generate three masks: scalp,
skull, and intracranial tissue. The scalp and skull masks were
merged with the above CSF, gray and white matter segmented
volume to generate a whole head model with 5 tissue types: scalp,
skull, CSF, gray matter, and white matter. Upon measurement,
the mean scalp thickness was found to be 11.463.0 mm (for
MNIz.220 mm; Figure 1A-inset), which is substantially at odds
with reported values [45,46,47,48]. This could arise due to various
factors related to the MRI acquisition and multi-scan averaging.
Hence, we eroded the scalp layer by 3 single-voxel steps using
Freesurfer’s mri_binarize tool, resulting in a mean scalp thickness
of 6.963.6 mm which is more in line with prior findings. We
proceeded with the 5-layer model with the eroded scalp
(Figure 1A).
For the Monte Carlo simulations, optical properties were
assigned to the five tissue types, per Table 1. These values
represent the mean of published optical properties across four
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 2 August 2013 | Volume 8 | Issue 8 | e66319
typically-used NIRS wavelengths: 690, 750, 780 and 830nm
[23,49,50]. For all tissue types, the tissue anisotropy parameter (g)
was set to 0.01–representing predominantly forward-scattered
light and the tissue index of refraction (n) was set to 1.
Monte Carlo Injection Points
Some 5,000 points were initially spaced 5 mm apart on a
95 mm radius sphere (the approximate radius of the Colin27
head). The sphere was centered within the Colin27 volume, and
the sphere points were moved radially toward or away from this
center voxel until they reached the surface of the Colin27 scalp.
This location list was then pruned to remove all points below the
cerebellum (MNIz,260 mm) as well as points that were both
below the eyes (MINz,235 mm) and anterior to the temporal
pole (MNIy.+35 mm). The resulting 3,555 points were spaced 3–
6 mm apart and covered all scalp regions overlying brain tissue
(Figure 1B). These points constituted the source locations and
directions for photon injection during Monte Carlo simulations.
Monte Carlo Simulations
We employed a three-dimensional Monte Carlo (MC) method
based on the tMCimg code described by Boas et al. [51], with the
general approach being described by Wang [52]. In brief, the
initial position and direction of a photon are defined as coming
from a point source with an initial survival weight W
0
set to 1. A
scattering length L is probabilistically calculated from an
exponential distribution, and the photon is moved through tissue
by this length. The photon’s weight, is incrementally decreased by
a factor of exp ({m
a
L) due to absorption, where m
a
is the
absorption coefficient of the tissue and Lis the length traveled by
the photon. A scattering angle is then calculated using the
probability distribution given by the Henyey-Greenstein phase
function, and a new scattering length is determined from an
exponential distribution. The photon is moved the new distance in
the updated direction defined by the scattering angle. This process
continued until the photon exited the medium or traveled longer
than 10ns, since the probability of photon detection in perfused
tissue after such a period of time is extremely small. When the
photon reaches a boundary, the probability of an internal
reflection is calculated from Fresnel’s formula. If a reflection
occurs, the photon is reflected back into the medium and
propagation continues with a new distance. Otherwise, the
migration of this photon is terminated and a new photon is
launched. The output of a given Monte Carlo simulation included
the photon fluence within the medium: an accumulation of all
photon weights within each voxel in the tissue, also known as the
2-point Green’s function. In addition, for each detector position
the MC simulation stores in a history file: (1) the number of
photons exiting within a 3 mm diameter of that detector location,
and (2) every photon’s path length traveled through each tissue
type that reached that detector. For each of the 3,555 simulations
in our study, 100 million (i.e., 10
8
) input photons were injected at
the source location. Individual MC simulations required 4–
5 hours of computation time and 1–2 GB of storage (uncom-
pressed).
Sensitivity Map Computation
Unlike x-rays which typically pass straight through biological
tissue, near-infrared photons scatter significantly as they travel
through tissue. The spatial probability distribution of photons
entering tissue at a source location, scattering through the tissue,
and being emitted at a particular detector location, defines the
spatial sensitivity profile (i.e., the ‘‘probed tissue’’) for that source-
detector (SD) pair. In linear DOI image reconstruction, this spatial
probability distribution is represented by a 3-point Green’s
function (see Eqn 1), which can be computed from two separate
2-point Green’s functions (i.e., 2 separate MC results), as described
below (cf. Figure 2).
The probability of a photon traveling from a point source,
~
rr
s
,to
any particular voxel inside the head,
~
rr, is represented by a single
Figure 1. Anatomy used for Monte Carlo simulations and processing. (A) Sagittal section (MNIx = 29 mm) through the segmented Colin27
head. Shades from white to dark gray are: scalp, skull, cerebrospinal fluid, white and gray matter, respectively. The inset shows the original, pre-
eroded scalp on the same slice. (B) Location and orientation of the 3,555 photon injection points around the Colin27 scalp used for the Monte Carlo
simulations. The injection vectors (yellow) are shown in relation to the scalp profile and underlying cortical surface (rendered brain). The nineteen
standard locations in the International 10–20 System are highlighted in blue. (C) Colorized shells representing the masks used for depth sensitivity
analysis (blue = scalp, green = skull, dark blue to pink = twenty-one ,2.8 mm thick shells).
doi:10.1371/journal.pone.0066319.g001
Table 1. Optical properties for scalp, skull, CSF, gray and
white matter used for all Monte Carlo simulations.
Tissue Type
m
a
(
mm
2
1
)
m
s
(
mm
2
1
)
Gray matter 0.019500 1.10
White matter 0.016900 1.35
CSF 0.002500 0.01
Skull 0.011925 0.92
Scalp 0.017275 0.72
doi:10.1371/journal.pone.0066319.t001
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 3 August 2013 | Volume 8 | Issue 8 | e66319
2-point Green’s function (W
0
in Eqn 1; see also Figure 2A,B). The
probability of a photon traveling from a point source, to any
particular voxel inside the head and then arriving a particular
detector location,
~
rr
d
, is represented by a 3-point Green’s function
[53], W (see Figure 2C,D). The 3-point Green’s function is
formally given by:
W
~
rr
s
,
~
rr
d
ðÞ~
ð
W
0
~
rr,
~
rr
s
ðÞG
~
rr
d
,
~
rrðÞd
3
r ð1Þ
where
~
rr
s
and
~
rr
d
are the locations of the source and detector,
respectively, and
~
rr is a position inside the head. W
0
is the fluence
provided by the MC 2-point file for the source point (e.g.
Figure 2B), and the function G is provided by a separate MC 2-
point file for the detector point. For continuous-wave NIRS
measurements, the 3-point function (e.g., Figure 2D) can be
generated by simply multiplying the 2-point function obtained
from the source location (e.g., Figure 2B) by the 2-point function
from the detector location, voxel by voxel. [For energy conserva-
tion, the data stored in each tMCimg 2-point file is first
normalized according to Eqn. (1) in [51].] Once computed, the
3-point function W represents the sensitivity of the optical
measurement for detecting changes at any point inside the
medium, or a ‘‘spatial sensitivity profile’’, from that SD pair
(Figure 2D) typically described as roughly banana-shaped. We
will refer to the sensitivity function W the ‘‘3- point sensitivity
function’’.
Spatial Sensitivity Analyses
To estimate NIRS sensitivity to our five different tissue types
within the head (e.g., Figure 1A), a given 3-point sensitivity map
was partitioned and accumulated. The total fluence change to a
perturbation in brain tissue, dw
total
, can be partitioned various
ways based on the 3-point maps. For example, one option is to
partition the total fluence change into two components: brain
fluence and non-brain fluence:
dw
total
~dw
brain
zdw
nonbrain
ð2Þ
If we assume the absorption change in brain tissue sampled by
that SD pair is uniform (and similarly for the non-brain tissue),
these can be represented by dm
brain
a
and dm
nonbrain
a
respectively. The
total fluence change can then be written in terms of the 3-point
functions W, partitioned into brain and non-brain components:
dw
total
~
X
voxels
W
brain
dm
brain
a
z
X
voxels
W
nonbrain
dm
nonbrain
a
ð3Þ
Since sensitivity is defined as the change in optical fluence
divided by change in m
a
(assuming scattering is constant), the
sensitivity to brain tissue and non-brain tissue can be derived by
taking the partial derivative of the above equation with respect to
m
brain
a
and m
nonbrain
a
, respectively:
Lw
total
Lm
brain
a
~
X
voxels
W
brain
ð4Þ
Lw
total
Lm
nonbrain
a
~
X
voxels
W
nonbrain
ð5Þ
The sensitivity to one partition (e.g., brain tissue) can therefore
be estimated by summing over all voxels in the 3-point weighting
function that comprise that partition,
X
W
brain
. This concept
generalizes from a brain/non-brain partition to any other
partitioning of the head. For example, one could separately sum
Figure 2. Photon propagation through scattering tissue. (A) Representation of a single photon moving through tissue, from the source, to an
arbitrary point inside the medium. Accumulation of photon weights during this process is the basis of a 2-point Green’s function. (B) Example 2-point
Green’s function, with colors representing the intensity of light reaching any given point in the tissue (truncated after a 5 order-of-magnitude
reduction in intensity from peak). (C) Representation of a single photon traveling from the source, to a point in the medium, and on to a detector; the
basis of a 3-point sensitivity function. (D) Example 3-point sensitivity function generated from two MC simulations (one for the source, one for the
detector) spaced 30 mm apart.
doi:10.1371/journal.pone.0066319.g002
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 4 August 2013 | Volume 8 | Issue 8 | e66319
over all gray matter voxels to compute the cumulative sensitivity to
gray matter, or sum over white matter voxels to compute the
sensitivity to white matter. One could also use geometric partition
partitions of the head, such as concentric ‘‘depth’’ shells described
further below.
Regional and global sensitivity analysis. To assess
regional NIRS sensitivity around the head, we performed
sensitivity analyses at the 19 standard points of the International
10–20 System in MNI coordinate space [54]. For each of the 19
locations, we first identified every pair of points from our 3,555
MCs that were separated by less than 60 mm and whose midpoint
lay within 10 mm of the given 10–20 location. We then generated
a 3-point map for each pair identified, and summed the 3-point
weights for each of the five tissue types to quantify the probability
of photons passing through each tissue type. This partitioning and
sensitivity computation was repeated for each pair in the list,
retaining SD separation information. Sensitivity and variability
was quantified by averaging results surrounding a given 10–20
location, in 5 mm bins. For example, to estimate sensitivity to gray
matter given a 30 mm separation, we averaged together the gray
matter partition sums for all SD pairs that were 27.5–32.5 mm
apart and also centered on a point within 10 mm of the 10–20
location of interest. This entire process was repeated for each SD
separation considered. Depending on the separation, anywhere
from 30–100 such SD pairs contributed to the average.
The result for a given 10–20 location was an average curve
representing the sensitivity and variability of a NIRS measurement
to each tissue type as a function of SD separation, centered on the
target location. The entire process was then repeated at each of the
nineteen 10–20 locations to quantify sensitivity and variability as a
function of position on the head. To estimate the mean global
sensitivity to the five tissue types, we averaged the curves from the
19 standard locations of the 10–20 system. In this analysis, we
dropped the error associated with individual SD-pairs surrounding
individual locations to highlight the variability across 10–20
location.
In practice, intensity contours such as those shown in Figure 2
are linearly related to the measurement of signal change. Optical
property changes occurring within the farthest contours from the
source and detector will produce smaller signal changes. Hence, a
more sensitive or lower-noise instrument will be required to detect
signals arising from those regions. The farthest contour to which
an instrument is sensitive, therefore, is characterized by the NIRS
instrument’s noise characteristics, sensitivity and dynamic range.
In our study, the entire sensitivity analysis process was performed
both before and after thresholding the 3-point functions at 5, 4, 3
or 2 orders of magnitude in sensitivity loss from peak (i.e., masking
at the 5th, 4th, 3rd or 2nd contour line in Figure 2). This simulates
instruments with progressively lower sensitivity, higher noise, and/
or a more restricted dynamic range.
Depth sensitivity. To assess NIN depth sensitivity, the
intracranial mask generated by FSL was eroded in three
dimensions 21 times via the mri_binarize tool (FreeSurfer,
v5.1), in two-voxel steps. Erosion in a voxellated space depends
upon the surface curvature, but the mean erosion step was
2.8 mm. Successive mask-pairs were then subtracted to generate
a series of twenty-one ‘‘shells’’, each on average 2.8 mm thick,
beginning at the inner edge of the skull and continuing toward
the center of the brain. These were then combined with the scalp
and skull masks to provide a separate head segmentation with 23
complete, non-overlapping shells (colors in Figure 1C). As with
the regional sensitivity analysis, the same groups of 3-point
sensitivity maps surrounding the nineteen 10–20 locations were
partitioned using each of these 23 shells (i.e., ignoring the 3
intracranial tissue types). Sensitivity weights were summed to
independently quantify cumulative photon fluence within each
layer in depth, irrespective of the five tissue types used in the MC
simulations themselves.
Results
Spatial Sensitivity for Near-Infrared Neuroimaging
A series of 3-point spatial sensitivity functions, W, are shown in
Figure 3 in a sagittal slice covering a wide range of SD separations.
Contour lines appear at each order of magnitude decrease in the
photon sensitivity profile from the peak voxel, and are truncated in
each figure after decaying 5 orders-of-magnitude. This represents
the approximate limits of a sensitive, low noise, and well-optimized
NIRS measurement; less sensitive devices or poorly optimized
dynamic range may only provide information up through the first
three or four contour lines.
Consistent with previous findings, as separations increase from 0
to 55 mm, the NIRS ‘‘banana’’ grows and extends deeper into the
brain. At a separation of 55 mm, the overlay shown intersects with
roughly the outer 17 mm of brain tissue. Careful visual inspection
of these examples also reveals that the relatively non-scattering
CSF layer distorted the normally smooth ovoid shape of the NIRS
banana that is found in a homogeneous medium [25]. Such
distortion is particularly noticeable when the CSF layer is more
than 1–2 millimeters thick and the CSF layer is near the edge of
the sensitivity profile (see the 5 and 10 mm separations in Figure 3).
These examples also suggest there is at least limited sensitivity to
gray matter even at SD separations less than 20 mm separations
that are typically assumed to provide zero brain sensitivity in an
adult human head.
Sensitivity as a Function of Source-Detector Separation
Figure 4 shows quantitatively the average proportion of
sensitivity to brain tissue (gray+white matter) and non-brain tissue
(scalp+skull+CSF) as a function of SD separation, along with
separate estimates for gray and white matter. The fluence mean
and variability within the indicated tissue type(s) was computed
across the nineteen 10–20 locations and error bars hence represent
an estimate of the variability over these 19 head locations. The
results were plotted as a function of SD separation. Two main
points should be noted. First, there was notable sensitivity to brain
tissue at a separation of 20 mm, representing approximately 6% of
the total sensitivity profile. (Smaller separations are not included
because there were too few source-detector pairs having a
midpoint of measurement within 10 mm of the 10–20 positions
for reliable estimation.) Second, at the largest SD separation
(65 mm), the estimated sensitivity to brain tissue reached
approximately 22% of the total sensitivity profile Thus, even at
these larger-than-typical separations, the NIRS measurement is
substantially more sensitive to the scalp, skull and CSF tissue
compartments than the gray and white matter components.
Figure 5 provides equivalent curves for scalp, skull and CSF
tissues.
The sensitivity to gray matter, of particular interest for NIN
measurements, was largely linear up through 45 mm separations,
at which point diminishing returns were observed for larger
separations. Linear regression across the 20 through 45 mm
separation values for gray matter revealed a slope of 0.0460.003
per centimeter. That is, for every 10 mm increase in SD
separation, sensitivity to gray matter increased an additional 4%
(from 6% at 20 mm, to 16% at 45 mm).
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 5 August 2013 | Volume 8 | Issue 8 | e66319
Figure 3. Photon sensitivity profile at a broad range of source-detector separations. Contours are drawn for each order of magnitude loss
in sensitivity from peak and are truncated after 5 orders of magnitude.
doi:10.1371/journal.pone.0066319.g003
Figure 4. Mean proportion of total sensitivity to the tissue types indicated as a function of source-detector separation. Errorbars
represent standard errors across all nineteen locations in the International 10–20 System. Separate curves represent pre-thresholding of the sensitivity
(3-point Green’s function) maps at 5, 4, 3, or 2 orders of magnitude (OM) reduction in sensitivity compared to peak, representing progressively less
optimal NIRS measurement systems. (A) Sensitivity to brain tissue = gray plus white matter. (B) Non-brain tissue = CSF plus skull plus scalp. (C) Gray
matter only. (D) White matter only.
doi:10.1371/journal.pone.0066319.g004
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 6 August 2013 | Volume 8 | Issue 8 | e66319
Sensitivity as a Function of Instrument Performance
The entire analysis described above was repeated after masking
the 3-point functions at 5 or (separately) 4, 3 or 2 orders of
magnitude sensitivity loss from peak (separate curves in each panel
in Figure 4 and Figure 5). The results with no masking and
masking at 5 orders of magnitude were indistinguishable and
hence only the latter is plotted for clarity. The analyses at 3 and 4
orders of magnitude, simulating progressively lower performing
NIRS instrumentation, provided qualitatively similar curves,
although absolute sensitivity was reduced (downward shifts) to
both gray and white matter layers as progressively more of the 3-
point function was masked. The sensitivity decrement in gray
matter was just significant (p,0.05) at most separations when
masking after 3 orders of magnitude, and reflected an 8–10%
decrease in sensitivity. The magnitude of this decrement was
roughly equivalent to the decrement associated with a 3 mm
increase in SD separation. Masking after only 2 orders of
magnitude sensitivity change approximately equivalent to an
instrument with 40 dB dynamic range exhibited substantial and
significant reductions in sensitivity, up to 50%, in all intracranial
tissue classes (CSF, gray and white matter).
NIRS Depth Sensitivity in the Brain
To estimate the NIRS sensitivity to intracranial tissue as a
function of depth in the Colin27 model we used the intracranial
shells depicted in Figure 1C and the described partitioning
method. Figure 6 plots the average sensitivity to each shell
starting with scalp, then skull, then each successive ,2.8-mm
intracranial shell averaged over all 10–20 positions. Note that
the first such shell (0–2.8 mm) typically intersected a substantial
portion of CSF plus a modest amount of gray matter. The next
intracranial shell (,2.8–5.6 mm below the inner skull surface)
typically contained substantially less CSF and considerably more
gray matter. The third intracranial shell (5.6–8.4 mm deep) begins
to have more contact with white matter voxels, and so on towards
the center of the brain (cf. Figure 1C).
In Figure 6, two primary effects are prominent. First, as SD
separation increases, the sensitivity to all layers except scalp also
increases (Figure 6A). For example, at a 20 mm separation, scalp
and skull provided approximately 84% and 15% of the NIRS
measurement sensitivity, whereas at 55 mm SD separations they
provided closer to 35% and 45% of the NIRS sensitivity. The
arbitrarily-selected 1% sensitivity line (y-axis = 10
22
) was exceeded
for approximately the outermost 11.2 mm of intracranial volume,
starting at a 25 mm SD separation.
The 3-point Greens function (Eqn. 1) is essentially exponential,
and hence we plotted log
10
(sensitivity) as a function of depth into
the intracranial space (Figure 6B). At any given SD separation,
sensitivity was indeed observed to decrease exponentially with
depth. Non-linear regression analysis was performed on the
sensitivity data (from the nineteen 10–20 positions) using an
exponential formula of the following form:
sensitivity~azb c
depth
ð6Þ
The regressions revealed significant b and c coefficients at each
separation, with models accounting for 53%–91% of the total
variance (see Table 2). As expected, the sensitivity at the innermost
edge of the skull increased significantly as SD separation increased
(T(11) = 16.2, p,0.0001; slope = ; adjusted R
2
= 0.96). In
addition, the exponential (decay) coefficient c increased signifi-
cantly with larger SD separations, reflecting a slower loss in
sensitivity with depth at larger SD separations. However, the c
coefficient plateaued at approximately a 40 mm separation
(Figure 7).
While depth sensitivity for a given SD separation can be
estimated directly from the coefficients in Table 2 plus Eqn. (6), we
also computed a ‘‘rule of thumb’’ formula. Averaging over the
coefficients for typically used NIRS separations of 20–40 mm
results in the following equation for the sensitivity in depth (S):
S ~0:075 0:85
depth
ð7Þ
Thus, at a ‘‘typical’’ separation, our MCs estimate a
0.075*0.85
0
= 0.075, or 7.5% sensitivity at the average inner
surface of the skull. At a depth 5 mm into the intracranial space,
the sensitivity would decrease, to 0.075*0.85
5
= 0.033 or 3.3%.
Such estimates are coarse, and assume overlying scalp and skull
layers comparable to those in our Colin27 template. However,
Equation (7) can provide a reasonable estimate of NIN sensitivity
as a function of depth. For comparison, quantitative depth
sensitivity measures at each 10–20 location for a typical SD
separation of 30 mm are listed in Table 3.
Discussion
Utilizing 3,555 Monte Carlo simulations, we generated
quantitative estimates of the spatial sensitivity and associated
variability of non-invasive NIRS to brain tissue in the Colin27
Figure 5. Mean proportion of total sensitivity to scalp, skull, and CSF as a function of source-detector separation. Errorbars represent
standard errors across all nineteen locations in the International 10–20 System. Separate curves again represent pre-thresholding of the sensitivity (3-
point Green’s function) maps at 5, 4, 3 or 2 orders of magnitude (OM) reduction in sensitivity compared to peak.
doi:10.1371/journal.pone.0066319.g005
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 7 August 2013 | Volume 8 | Issue 8 | e66319
brain. Sensitivity as a function of source-detector separation was
qualitatively consistent with previous reports [23,24,31,32,36].
Global sensitivity to brain tissue reached 20% (and gray
matter = 17%) of the total sensitivity at a SD separation of 55
mm, with the remaining sensitivity coming from scalp, skull and
CSF. More typical SD separations of 25–35 mm exhibited brain
sensitivities in the range of 8–13%. As expected, sensitivity was
degraded when we simulated under-performing NIRS instrumen-
tation, with a typical sensitivity reduction equivalent to the
decrease expected from a 2.5 mm increase in SD separation.
Performance appeared to drop quickly with more severely
degraded instruments (as dynamic range fell from ,60 dB to
,40 dB). Finally, the sensitivity in depth through the intracranial
space was found to be exponentially decreasing, with detailed
exponential coefficients given in Table 2, and an approximation
for sensitivity-in-depth given by Eqn. (7). NIN signals thus appear
strongly biased to absorption changes occurring in the outer 10–
15 mm of the intracranial space. While these results were based on
the thinned-scalp version of Colin27, we also performed a full set
(3,555) of MC simulations on the original, thicker scalp.
Qualitatively, we replicated all results reported herein, with the
primary difference being uniformly but modestly lower sensitivity
to brain tissue in the thicker-scalp model.
Sensitivity versus SD Separation
While an increase in brain sensitivity was anticipated with larger
SD separations [31,32,35,40], the magnitude and variability of this
relationship around the head, and in a realistic geometry, were not
well characterized. Our MC results suggest that, globally, as much
as 6% of the NIRS sensitivity comes from brain tissue at SD
separations of 20 mm. This probability rose sigmoidally over the
range of separations we investigated, increasing to 20% of the
NIRS sensitivity coming from brain tissue at a 55 mm separation.
This is qualitatively consistent with recent studies that examined
only small regions in different head models [35,39,40].
Our results did quantitatively differ to some degree from this
prior work. For example, our observed sensitivity of ,10% at a
separation of 30–35 mm was slightly higher than that found when
using the Chinese head dataset [55], for which 8% was reported at
the same 30–35 mm separation [39]. However, that study as
well as the others that considered realistic head models only
examined very small portions of a different head model, whereas
our mean and variability estimates apply to the entire Colin27
head.
Variability in NIRS sensitivity to brain had not been previously
assessed. In our study, variability in sensitivity was modest within
regions, ranging from 10–18%. Variability was somewhat more
substantial between regions, with standard errors across the 10–20
locations spanning 20% of the mean sensitivity values. The
variability around the head, therefore, appears to be an important
contributor to NIN sensitivity. This variability is presumed to be
related to the different tissue layering found across different head
regions.
Figure 6. Mean NIRS depth sensitivity in the brain plotted in two orthogonal ways, by SD separation. (A) The top two traces represent
scalp (blue) and skull (green) sensitivity. Sensitivity to scalp and skull were equal at a SD separation of 25 mm. On average, 1% or more of the
sensitivity profile was achieved for all of the most superficial 11.2 mm of the intracranial volume at SD separations of 25 mm or greater (circle). (B)
Intracranial sensitivity in depth as a function of source-detector separation (excluding scalp and skull). At all separations, sensitivity decreases
exponentially with depth (i.e., linear curves through ,15 mm depth on this semilog plot).
doi:10.1371/journal.pone.0066319.g006
Figure 7. Fitted exponential decay coefficient, c, from the
sensitivity function in Eqn. (6) as a function of SD separation.
The asymptote at ,40 mm separations means that further increasing
the SD separation provides diminishing returns for NIRS sensitivity to
brain function.
doi:10.1371/journal.pone.0066319.g007
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 8 August 2013 | Volume 8 | Issue 8 | e66319
There was a notable gain in sensitivity with SD separation,
particularly at greater than 25 mm separations. In one prior study,
it was recommended that NIRS researchers avoid SD separations
greater than 30–35 mm [39]. While this does not appear to be
supported by our findings, it is important to note that our estimates
do not consider the achievable signal-to-noise ratio (SNR). If an
instrument cannot detect a reliable signal at 40 or 50 mm, larger
separations will not improve brain sensitivity. A trade-off therefore
needs to be made between greater sensitivity to deeper layers and
the resulting loss in SNR. All NIRS instruments have a limit on the
separation where reliable SNR can be achieved, which is readily
determined by experimentation. We investigated a broad range of
separations, spanning the typical capabilities of current instru-
ments, to help individual investigators determine this trade-off for
specific devices.
Based on our results, therefore, NIRS probes for non-invasive
neuromonitoring in adults should ideally be designed with 30–
35 mm SD separations, or larger, assuming the instrument can
Table 2. Regression results for sensitivity as a function of depth.
SD Model Coef. Std. 95% Conf. Interval N Sensitivity at Depth (mm)
Sep. Param Value Err. T p Low High Adj. R2 0 5 10
5 a 0.0000 0.0001 20.6 0.550 20.0002 0.0001 378
b 0.0058 0.0003 19.3 0.000 0.0052 0.0064 0.526 0.58% 0.19% 0.06%
c 0.8042 0.0185 43.4 0.000 0.7678 0.8406
10 a 20.0001 0.0001 20.7 0.491 20.0004 0.0002 399
b 0.0133 0.0005 25.3 0.000 0.0123 0.0143 0.647 1.32% 0.48% 0.17%
c 0.8176 0.0132 62.1 0.000 0.7917 0.8435
15 a 20.0002 0.0003 20.7 0.458 20.0007 0.0003 399
b 0.0261 0.0009 27.8 0.000 0.0243 0.0280 0.687 2.59% 1.00% 0.38%
c 0.8292 0.0113 73.6 0.000 0.8071 0.8514
20 a 20.0003 0.0004 20.9 0.392 20.0011 0.0004 399
b 0.0431 0.0013 32.7 0.000 0.0405 0.0457 0.753 4.27% 1.74% 0.70%
c 0.8377 0.0091 91.8 0.000 0.8198 0.8557
25 a 20.0005 0.0005 21.0 0.341 20.0015 0.0005 399
b 0.0614 0.0017 36.5 0.000 0.0581 0.0647 0.793 6.09% 2.59% 1.09%
c 0.8448 0.0078 107.8 0.000 0.8294 0.8602
30 a 20.0006 0.0006 21.1 0.276 20.0018 0.0005 399
b 0.0781 0.0019 41.6 0.000 0.0744 0.0818 0.832 7.74% 3.37% 1.44%
c 0.8483 0.0067 125.7 0.000 0.8350 0.8616
35 a 20.0008 0.0006 21.2 0.221 20.0020 0.0005 399
b 0.0913 0.0020 45.8 0.000 0.0873 0.0952 0.857 9.05% 4.01% 1.75%
c 0.8514 0.0060 141.5 0.000 0.8395 0.8632
40 a 20.0009 0.0006 21.3 0.182 20.0021 0.0004 399
b 0.1018 0.0021 49.3 0.000 0.0977 0.1059 0.875 10.09% 4.51% 1.99%
c 0.8528 0.0055 154.1 0.000 0.8419 0.8637
45 a 20.0009 0.0007 21.4 0.165 20.0022 0.0004 399
b 0.1095 0.0021 51.5 0.000 0.1054 0.1137 0.883 10.86% 4.89% 2.17%
c 0.8542 0.0053 162.3 0.000 0.8438 0.8645
50 a 20.0010 0.0007 21.4 0.163 20.0023 0.0004 399
b 0.1157 0.0022 52.7 0.000 0.1114 0.1200 0.889 11.47% 5.20% 2.33%
c 0.8554 0.0051 167.9 0.000 0.8454 0.8655
55 a 20.0010 0.0007 21.5 0.134 20.0024 0.0003 378
b 0.1231 0.0022 54.8 0.000 0.1187 0.1276 0.901 12.21% 5.45% 2.40%
c 0.8528 0.0050 171.2 0.000 0.8430 0.8626
60 a 20.0011 0.0007 21.5 0.132 20.0025 0.0003 378
b 0.1264 0.0023 56.0 0.000 0.1219 0.1308 0.905 12.53% 5.65% 2.52%
c 0.8545 0.0048 176.9 0.000 0.8450 0.8640
65 a 20.0011 0.0007 21.6 0.112 20.0025 0.0003 357
b 0.1321 0.0023 57.6 0.000 0.1275 0.1366 0.914 13.09% 5.81% 2.54%
c 0.8518 0.0048 178.6 0.000 0.8425 0.8612
doi:10.1371/journal.pone.0066319.t002
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 9 August 2013 | Volume 8 | Issue 8 | e66319
provide adequate SNR at those separations. Note that this
recommendation should also be weighed against spatial resolution
when choosing SD separations. As SD separations increase, the
effective resolution of the NIRS measurement decreases. In some
studies, the need for a higher spatial resolution (and hence shorter
SD separations) may outweigh the need for more or deeper
sensitivity (and hence larger SD separations).
For close SD pairs, separations of 20 mm included on average
6% of the photon fluence (sensitivity) attributable to brain tissue.
Thus, the commonly used rule of thumb namely, that
measurements with 20 mm or smaller separations are minimally
sensitive to the brain in adult humans may need to be revised.
Clearly, brain activation that is both strong and superficial could
still be detected with a 20 mm separation. As mentioned earlier,
this would be even more prominent if superficial layers exhibit
little or no change in optical properties, or such changes are
eliminated or suitably accounted for. Since there is substantial
vascularization of and hemodynamic activity in scalp, skull and
pial vessels, experimental manipulations to eliminate such signals
may be difficult. When seeking to independently measure
Table 3. Estimated relative NIRS sensitivity (proportions) as a function of depth for a SD separation of 30 mm.
Shell Depth Range (mm)
Location N Scalp Skull 0–2.8 2.8–5.6 5.6–8.4 8.4–11.2 11.2–14 14–16.8 16.8–19.6
Fp1 45 0.16 0.70 0.059 0.046 0.0202 0.0078 0.00272 0.00079 0.00020
(0.03) (0.02) (0.009) (0.007) (0.0032) (0.0013) (0.00051) (0.00017) (0.00006)
Fp2 47 0.21 0.64 0.064 0.047 0.0190 0.0069 0.00241 0.00073 0.00019
(0.04) (0.03) (0.011) (0.008) (0.0039) (0.0016) (0.00062) (0.00022) (0.00008)
Fz 66 0.26 0.52 0.047 0.046 0.0406 0.0302 0.02048 0.01339 0.00762
(0.02) (0.02) (0.004) (0.004) (0.0032) (0.0033) (0.00307) (0.00255) (0.00187)
F3 78 0.15 0.64 0.084 0.071 0.0341 0.0145 0.00591 0.00217 0.00071
(0.01) (0.01) (0.010) (0.007) (0.0030) (0.0016) (0.00080) (0.00035) (0.00013)
F4 91 0.17 0.64 0.085 0.066 0.0273 0.0097 0.00341 0.00111 0.00032
(0.01) (0.01) (0.008) (0.006) (0.0024) (0.0010) (0.00046) (0.00019) (0.00007)
F7 76 0.65 0.21 0.055 0.046 0.0218 0.0088 0.00318 0.00111 0.00040
(0.06) (0.03) (0.012) (0.009) (0.0041) (0.0016) (0.00062) (0.00025) (0.00012)
F8 71 0.69 0.20 0.041 0.036 0.0185 0.0077 0.00291 0.00114 0.00048
(0.05) (0.03) (0.008) (0.007) (0.0033) (0.0015) (0.00065) (0.00029) (0.00015)
C3 90 0.34 0.41 0.068 0.069 0.0520 0.0297 0.01501 0.00713 0.00325
(0.02) (0.01) (0.005) (0.005) (0.0044) (0.0029) (0.00169) (0.00087) (0.00043)
C4 83 0.36 0.46 0.052 0.053 0.0389 0.0220 0.01047 0.00467 0.00201
(0.02) (0.01) (0.005) (0.005) (0.0037) (0.0023) (0.00126) (0.00064) (0.00031)
Cz 43 0.41 0.30 0.031 0.034 0.0350 0.0358 0.03275 0.02818 0.02398
(0.02) (0.02) (0.002) (0.002) (0.0016) (0.0014) (0.00168) (0.00159) (0.00132)
P3 66 0.26 0.52 0.087 0.073 0.0359 0.0146 0.00566 0.00210 0.00070
(0.02) (0.01) (0.009) (0.007) (0.0044) (0.0023) (0.00107) (0.00047) (0.00019)
P4 65 0.25 0.49 0.095 0.081 0.0452 0.0209 0.00901 0.00346 0.00118
(0.02) (0.01) (0.007) (0.005) (0.0038) (0.0024) (0.00129) (0.00061) (0.00025)
Pz 38 0.48 0.34 0.037 0.037 0.0337 0.0253 0.01693 0.01081 0.00683
(0.06) (0.03) (0.007) (0.007) (0.0064) (0.0051) (0.00369) (0.00245) (0.00159)
O1 34 0.17 0.56 0.132 0.093 0.0333 0.0103 0.00303 0.00084 0.00021
(0.01) (0.01) (0.007) (0.005) (0.0027) (0.0011) (0.00039) (0.00013) (0.00004)
O2 43 0.19 0.51 0.133 0.096 0.0400 0.0147 0.00528 0.00192 0.00065
(0.01) (0.02) (0.009) (0.005) (0.0032) (0.0020) (0.00111) (0.00056) (0.00024)
T3 80 0.68 0.12 0.083 0.066 0.0304 0.0112 0.00387 0.00127 0.00038
(0.02) (0.01) (0.008) (0.006) (0.0029) (0.0011) (0.00043) (0.00016) (0.00006)
T4 66 0.70 0.13 0.076 0.057 0.0264 0.0103 0.00371 0.00123 0.00037
(0.03) (0.01) (0.009) (0.006) (0.0030) (0.0012) (0.00046) (0.00016) (0.00006)
T5 62 0.35 0.49 0.078 0.052 0.0169 0.0048 0.00134 0.00035 0.00006
(0.03) (0.02) (0.012) (0.008) (0.0030) (0.0010) (0.00034) (0.00013) (0.00004)
T6 63 0.40 0.47 0.074 0.043 0.0131 0.0036 0.00094 0.00021 0.00002
(0.05) (0.03) (0.011) (0.007) (0.0025) (0.0008) (0.00024) (0.00007) (0.00002)
doi:10.1371/journal.pone.0066319.t003
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 10 August 2013 | Volume 8 | Issue 8 | e66319
peripheral tissues like scalp and skull, as when making multi-
distance measurements to correct for superficial layer hemody-
namics [56,57,58,59,60], SD separations shorter than 10 mm
should be strongly considered.
Sensitivity and NIRS Instrument Performance
When comparing the different curves within panels in Figure 4
and Figure 5, the peak sensitivity dropped only about 10% when
assuming a NIRS instrument with a lower dynamic range or
optimization: for gray matter at a separation of 55 mm, the
reduction was from 0.20 to 0.18. This dropped more precipitously
(by up to 50% of the original value) to 0.10 with poor dynamic
range or optimization. This suggests the importance of optimizing
instrument sensitivity and reducing all types of noise and signal
interference when making NIN measurements [57,60,61]. This
appears to be particularly for instruments with inherently modest
dynamic range (,60 dB).
Brain Sensitivity in Depth
Our shell-based depth analysis revealed exponentially decreasing
sensitivity with depth at all SD spacings, with expectedly different
intercepts at each separation. Upon reaching 11.2 mm into the
brain, relative sensitivity greater than 1% was only found with
25 mm or greater separations. Thus, in our Colin27 head model,
substantial separation appears required to achieve even modest
sensitivity past the outer 10 mm of intracranial space. Two
important qualifiers should be noted. First, the sensitivity beyond
10 mm into the intracranial space is strongly limited by the
thickness of the overlying tissue layers, including scalp and skull. In
our model these layers together averaged ,13 mm thick. Other
individuals with thicker or thinner scalp and skull layers would be
expected to have decreased or increased overall sensitivities,
respectively. While these are not controllable parameters, they
should be considered with respect to study design. Second, the
ability to detect absorption changes is also a function of contrast and
noise. A large enough change in optical properties (i.e., absorption
or scattering), a large enough region of change, or a low enough
noise floor (including noise generated from systemic physiological
processes as discussed above), could enable detection of events
occurring substantially deeper than 10 mm. Mathematically, the
detectability of a specific functional event depends on the contrast-
to-noise ratio. Usually one uses CNR = 1 (contrast is equal to the
noise floor) to define the minimum detectable contrast. Hence, we
have the equation contrast~
X
W dm
a
~noise . We discussed
at length the role of
X
W in this paper. However, dm
a
(the change
in optical properties) is clearly equally important, such that a
doubling of absorption coefficients would lead to a doubling of the
detected contrast.
In terms of noise, if there was precisely zero change in optical
properties in overlying layers (e.g., scalp and skull), or the changes
in overlying layers can be otherwise accounted for, then any
detected change must have come from deeper layers, regardless of
the relative sensitivity at that depth. The magnitude of optical
contrast or the level of noise suppression required for detection of
brain activity at, say, a 20 mm depth into the intracranial space
remains to be investigated.
Study Limitations
Hair is invisible to MRI, and hence was not considered in this
study. Hair can mechanically interfere with NIN measurements
and also has its own absorption spectrum, providing two separate
reductions in NIRS sensitivity. However, in the absence of motion,
hair is expected to be a substantially less dynamic absorber than
scalp or skull tissue, as it lacks hemodynamic and most other time-
dependent physiologic processes. Thus, while the sensitivity in
regions with hair will generally be lower than in regions without
hair (all else being equal), the reductions are expected to be
essentially fixed for a given hair color and density. To our
knowledge, no quantification of the effect of hair on NIN
measurements has yet been performed. Also, the specific methods
used to segment the Colin template may have led to over- or
under-estimates of tissue thicknesses or volumes. Proper assess-
ment of this hypothesis requires re-segmentation and re-running
all Monte Carlo simulations and follow-on analyses and hence was
beyond the scope of this study. Within the segmented regions, our
MC simulations used average optical properties, largely to reduce
the number of separate MCs that had to be performed. While we
have performed a few MC runs using wavelength-specific optical
properties and only observed subtle differences in sensitivity
profiles, systematic quantification or comparison across wave-
lengths in realistic head geometries remains as future work.
Finally, while our MC simulations covered the Colin27 head
model comprehensively, we only examined a single head model.
Different individuals will exhibit different scalp and skull
thicknesses, CSF distributions, cortical folding patterns, white
matter composition, and so forth. For example, females have been
found to have different skull thickness distributions than males
[62,63], children have overall thinner scalps and skulls than our
model [64], older individuals may have different baseline
scattering properties [65], and individuals vary in terms of the
amount and distribution of CSF [66]. Direct assessments of each
of these parameters on NIN sensitivity remain to be performed.
However, our large set of MC simulations did encompass a
substantial range of scalp, skull and CSF combinations, and our
associated variability estimates provide a guide to the importance
of some of these issues.
Conclusions
This study represents the most comprehensive characterization
conducted to date of NIRS sensitivity to brain tissue, including the
importance of source-detector separations, variability in NIRS
sensitivity around the head, the influence of NIRS instrument
performance, as well as NIN sensitivity in depth. Our results
suggest that increasing the source-detector separation past 20 mm
monotonically increases sensitivity to brain tissue, and hence the
larger the separation the better in terms of brain sensitivity.
However, diminishing returns appear to begin around 40–50 mm
SD separations, and sensitivity must also be balanced against the
SNR that can be achieved with any particular instrument at large
separations, as well as the spatial resolution required. Our MC
simulations further suggest that, while the depth sensitivity of
NIRS is not strictly limited, NIRS sensitivity decreases exponen-
tially with depth into the intracranial space and hence NIN signals
are strongly biased towards the outermost 10–15 mm of the
intracranial volume. The detailed quantitative information pro-
vided here can help investigators better design and plan
experiments, head probes and instruments for making NIRS
measurements, as well as providing guidance when interpreting
NIRS studies in terms of the likely sources of the observed signals.
Acknowledgments
We wish to thank our anonymous reviewers for their assistance in clarifying
and enhancing this work.
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 11 August 2013 | Volume 8 | Issue 8 | e66319
Author Contributions
Conceived and designed the experiments: GS ZL QZ. Performed the
experiments: GS ZL QZ. Analyzed the data: GS ZL QZ. Contributed
reagents/materials/analysis tools: GS ZL QZ. Wrote the paper: GS ZL
QZ.
References
1. Villringer A, Chance B (1997) Non-invasive optical spectroscopy and imaging of
human brain function. Trends Neurosci 20: 435–442.
2. Strangman G, Boas DA, Sutton JP (2002) Non-invasive neuroimaging using
near-infrared light. Biol Psychiatry 52: 679–693.
3. Ferrari M, Mottola L, Quaresima V (2004) Principles, techniques, and
limitations of near infrared spectroscopy. Can J Appl Physiol 29: 463–487.
4. Pogue BW, Patterso n MS (1994 ) F requency-domain optical a bsorption
spectroscopy of finite tissue volumes using diffusion theory. Physics in medicine
and biology 39: 25.
5. Moulton JD (1990) Diffusion modelling of picosecond laser pulse propagation in
turbid media: McMaster University.
6. Ishimaru A (1989) Diffusion of light in turbid material. Appl Opt 28: 6.
7. Profio AE (1989) Light transport in tissue. Appl Opt 28: 7.
8. Arridge S, Cope M, Delpy DT (1992) The theoretical basis for the determination
of optical pathlengths in tissue: temporal and frequency analysis. Phys Med Biol
37: 30.
9. Patterson MS, Chance B, Wilson BC (1989) Time resolved reflectance and
transmittance for the non-invasive measurement of tissue optical properties.
Appl Opt 28: 7.
10. Haselgrove JC, Schotland JC, Leigh JS (1992) Long-time behavior of photon
diffusion in an absorbing medium: application to time-resolved spectroscopy.
Appl Opt 31: 6.
11. Schweiger M, Arridge S, Delpy DT (1993) Application of the finite-element
method for the forward and inverse models in optical tomography. Journal of
Mathematical Imaging and Vision 3: 263–283.
12. Arridge S, Schweiger M, Hiraoka M, Delpy DT (1993) A finite element
approach for modeling photon transport in tissue. Proc SPIE 1888: 12.
13. Cui WJ, Ostrander LE (1992) The relationship of surface reflectance
measurements to optical properties of layered biologica l media. IEEE Trans
Biomed Eng 39: 194–201.
14. Hielscher AH, Alcouffe RE, Barbour RL (1998) Comparison of finite-difference
transport and diffusion calculations for photon migration in homogeneous and
heterogeneous tissues. Phys Med Biol 43: 1285–1302.
15. Arridge SR, Schweiger M, Hiraoka M, Delpy DT (1993) A finite element
approach for modeling photon transport in tissue. MedPhys 20: 299–309.
16. Jacques SL, Wang L (1995) Monte Carlo modeling of light transport in tissues.
In: Welch AJ, van Gemert JC, editors. Optical-Response of Laser-Irradiated
Tissue. New York: Plenum. 73–100.
17. Ren K, Abdoulaev GS, Bal G, Hielscher AH (2004) Algorithm for solving the
equation of radiative transfer in the frequency domain. Opt Lett 29: 578–580.
18. Wang L, Jacques SL (1993) Hybrid model of Monte Carlo simulation and
diffusion theory for light reflectance by turbid media. J Opt Soc Am A Opt
Image Sci Vis 10: 1746–1752.
19. Wang L, Jacques SL (1994) Optimized radial and angular positions in Monte
Carlo modeling. Med Phys 21: 1081–1083.
20. Wang L, Jacques SL, Zheng L (1995) MCML–Monte Carlo modeling of light
transport in multi-layered tissues. Comput Methods Programs Biomed 47: 131–
146.
21. Van der Zee PF, Delpy DT (1987) Simulation of the point spread function for
light in tissue by a Monte Carlo method. Adv Exp Med Biol 215: 18.
22. Wilson MA, McNaughton BL (1993) Dynamics of the hippocampal ensemble
code for space. Science 261: 1055–1058.
23. Okada E, Firbank M, Schweiger M, Arridge SR, Cope M, et al. (1997)
Theoretical and experimental investigation of near-infrar ed light propag ation in
a model of the adult head. Appl Opt 36: 21–31.
24. Hayashi T, Kashio Y, Okada E (2003) Hybrid Monte Carlo-diffusion method
for light propagation in tissue with a low-scattering region. Appl Opt 42: 2888–
2896.
25. Okada E, Delpy DT (2003) Near-infrared light propagation in an adult head
model. II. Effect of superficial tissue thickness on the sensitivity of the near-
infrared spectroscopy signal. Appl Opt 42: 2915–2922.
26. Okada E, Delpy DT (2003) Near-infrared light propagation in an adult head
model. I. Modeling of low-level scattering in the cerebrospinal fluid layer. Appl
Opt 42: 2906–2914.
27. Kawaguchi H, Hayashi T, Kato T, Okada E (2004) Theoretical evaluation of
accuracy in position and size of brain activity obtained by near-infrared
topography. Phys Med Biol 49: 2753–2765.
28. Okada E, Firbank M, Delpy DT (1995) The effect of overlying tissue on the
spatial sensitivity profile of near-infrared spectroscopy. Phys Med Biol 40: 2093–
2108.
29. Okada E (2000) The effect of superficial tissue of the head on spatial sensitivity
profiles for near infrared spectroscopy and imaging. Optical Review 7: 375–382.
30. Okada E, Firbank M, Schweiger M, Arridge SR, Hebden JC, et al. (2005)
Experimental measurements on phantoms and Monte Carlo simulation to
evaluate the effect of inhomogeneity on optical pathlength. Proc SPIE 2389.
31. Hiraoka M, Firbank M, Essenpreis M, Cope M, Arridge SR, et al. (1993) A
Monte Carlo investigation of optical pathlength in inhomogeneous tissue and its
application to near-infrared spectroscopy. Phys Med Biol 38: 1859–1876.
32. Fukui Y, Ajichi Y, Okada E (2003) Monte Carlo prediction of near-infrared light
propagation in realistic adult and neonatal head models. Appl Opt 42: 2881–
2887.
33. Zhang Q, Strangman GE (2007) Electro-optical System, Apparatus and Method
for Ambulatory Monitoring. USA: The General Hospital Corporation.
34. Hoshi Y, Shimada M, Sato C, Iguchi Y (2005) Reevaluation of near-infrared
light propagation in the adult human head: implications for functional near-
infrared spectroscopy. J Biomed Opt 10: 064032.
35. Mansouri C, L’Huillier JP, Kashou NH, Humeau A (2010) Depth sensitivity
analysis of functional near-infrared spectroscopy measurement using three-
dimensional Monte Carlo modelling-based magnetic resonance imaging. Lasers
Med Sci 25: 431–438.
36. Hayashi T, Kashio Y, Okada E (2003) Effect of Superficial Tissue on Spatial
Sensitivity Profile of Optical Brain Activity Measurement Evaluated by Direct
Hybrid Monte Carlo-Diffusion Method. Optical Review 10: 5.
37. Firbank M, Arridge SR, Schweiger M, Delpy DT (1996) An investigation of light
transport through scattering bodies with non-scattering regions. Phys Med Biol
41: 767–783.
38. Heiskala J, Neuvonen T, Grant PE, Nissila I (2007) Significance of tissue
anisotropy in optical tomography of the infant brain. Appl Opt 46: 1633–1640.
39. Li T, Gong H, Luo Q (2011) Visualization of light propagation in visible
Chinese human head for functional near-infrared spectroscopy. J Biomed Opt
16: 045001.
40. Custo A, Wells WM 3rd, Barnett AH, Hillman EM, Boas DA (2006) Effective
scattering coefficient of the cerebral spinal fluid in adult head models for diffuse
optical imaging. Appl Opt 45: 4747–4755.
41. Holmes CJ, Hoge R , Collins L, Woods R, Toga AW, et al. (1998) Enhancement
of MR images using registration for signal averaging. J Comput Assist Tomogr
22: 10.
42. Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, et al.
(2004) Advances in f unctional and structural MR image analysis and
implementation as FSL. Neuroimage 23 Suppl 1: S208–219.
43. Ashburner J, Friston KJ (2005) Unified segmentation. NeuroImage 26: 13.
44. Smith SM (2002) Fast robust automated brain extraction. Hum Brai n Mapp 17:
143–155.
45. Hori H, Moretti G, Rebora A, C rovato F (1972) The thickness of human scalp:
normal and bald. J Invest Dermatol 58: 396–399.
46. Oldendorf WH, Iisaka Y (1969) Interference of scalp and skull with external
measurements of brain isotope content. 1. Isotope content of scalp and skull.
J Nucl Med 10: 177–183.
47. Todd TW, Kuenzel W (1924) The Thick-of the Scalp. J Anat 58: 231–249.
48. Babiloni F, Babiloni C, Carducci F, Del Gaudio M, Onorati P, et al. (1997) A
high resolution EEG method based on the correction of the surface Laplacian
estimate for the subject’s variable scalp thickness. Electroencephalogr Clin
Neurophysiol 103: 486–492.
49. Bevilacqua F, Pigues D, Marquet P, Gross JD, Tromberg BJ, et al. (1999) In vivo
local determination of tissue optical properties: applications to human brain.
Appl Opt 38: 12.
50. Torricelli A, Pifferi A, Taroni P, Giambattistelli E, Cubeddu R (2001) In vivo
optical characterization of human tissues from 610 to 1010 nm by time-resolved
reflectance spectroscopy. Phys Med Biol 46: 11.
51. Boas D, Culver J, Stott J, Dunn A (2002) Three dimensional Monte Carlo code
for photon migration through complex heterogeneous media including the adult
human head. Opt Express 10: 159–170.
52. Wang L, Jacques SL, Zheng L (1995) MCML–Monte Carlo modeling of light
transport in multi-layered tissues. Comput Methods Programs Biomed 47: 16.
53. Yodh A, Chance B (1995) Spectroscopy and imaging with diffusing light. Physics
Today 48: 34–40.
54. Okamoto M, Dan H, Sakamoto K, Takeo K, Shimizu K, et al. (2004) Three-
dimensional probabilistic anatomical cranio-cerebral correlation via the
international 10–20 system oriented for transcranial functional brain mapping.
Neuroimage 21: 99–111.
55. Zhang SX, Heng PA, Liu ZJ, Tan LW, Qiu MG, et al. (2004) The Chinese
Visible Human (CVH) datasets incorporate technical and imaging advances on
earlier digital humans. J Anat 204: 165–173.
56. Hueber D, Franceschini M, Ma H, Xu Q, Ballesteros J, et al. (2001) Non-
invasive and quantitative near-infrared hemoglobin spectrometry in the piglet
brain during hypoxic stress, using a frequency-domain multi-distance instru-
ment. Phys Med Biol 46: 41–62.
57. Zhang Q, Strangman GE, Ganis G (2009) Adaptive filtering to reduce global
interference in non-invasive NIRS measures of brain activation: how well and
when does it work? NeuroImage 45: 788–794.
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 12 August 2013 | Volume 8 | Issue 8 | e66319
58. Zhang Q, Brown EN, Strangm an GE (2007) Adaptive filtering to reduce global
interference in evoked brain activity detection: a human subject case study.
J Biomed Opt 12: 064009.
59. Zhang Q, Brown EN, Strangman GE (2007) Adaptive filtering for global
interference cancellation and real-time recovery of evoked brain activity: a
Monte Carlo simulation study. J Biomed Opt 12: 044014.
60. Saager R, Berger A (2008) Measurement of layer-like hemodynamic trends in
scalp and cortex: implications for physiological baseline suppression in functional
near-infrared spectroscopy. J Biomed Opt 13: 034017.
61. Gratton G, Corballis PM (1995) Removing the heart from the brain:
compensation for the pulse artifact in the photon migration signal. Psychophys-
iology 32: 292–299.
62. Li H, Ruan J, Xie Z, Wang H, Liu W (2007) Investigation of the critical
geometric characteristics of living human skulls utilising medical image analysis
techniques. International Journal of Vehicle Safety 2: 345–367.
63. Law SK (1993) Thickness and resistivity variations over the upper surface of the
human skull. Brain Topogr 6: 99–109.
64. Adeloye A, Kattan KR, Silverman FN (1975) Thickness of the normal skull in
the Am erican Blacks and Whites. Am J Phys Anthropol 43: 23–30.
65. Duncan A, Meek JH, Clemence M, Elwell CE, Fallon P, et al. (1996)
Measurement of cranial optical path length as a function of age using phase
resolved near infrared spectroscopy. Pediatr Res 39: 889–894.
66. Hasan KM, Sankar A, Halphen C, Kramer LA, Brandt ME, et al. (2007)
Development and organization of the human brain tissue compartments across
the lifespan using diffusion tensor imaging. Neuroreport 18: 1735–1739.
NIRS Depth and Spatial Sensitivity
PLOS ONE | www.plosone.org 13 August 2013 | Volume 8 | Issue 8 | e66319